Development and Validation of Stability-Indicating RP-HPLC Method for Determination of Atazanavir and Ritonavir
Padmanabh B. Deshpande*, Santosh R. Butle
Department of Pharmaceutical Chemistry, School of Pharmacy, Swami Ramanand Teerth Marathwada University, Nanded-431 606
*Corresponding Author E-mail: padmanabh77@yahoo.co.in
ABSTRACT:
A new simple, accurate, precise and selective stability- indicating high performance liquid chromatographic (HPLC) method has been developed and validated for simultaneous estimation of Atazanavir and Ritonavir in tablet dosage form. An isocratic, reverse phase HPLC method was developed on Jasco HPLC system equipped with Grace C18 column (150 x 4.6 mm i.d.) using 10 mM Potassium dihydrogen phosphate buffer: Acetonitrile: Methanol (40: 40: 20, v/ v/v) as mobile phase and detection was carried out at 240 nm. Results were linear in the range of 50-500 µg mL-1 for Atazanavir and 25-400 µg mL-1 for Ritonavir. The retention time for Atazanavir and Ritonavir were 5.54 ± 0.02 and 6.82 ± 0.07 min. respectively. The drugs were subjected to stress condition of hydrolysis (acid, base), oxidation, photolysis and thermal degradation.
KEYWORDS: RP-HPLC, Atazanavir, Ritonavir, Forced degradation.
INTRODUCTION:
Atazanavir (ATV), chemically, (3S,8S,9S,12S)-3,12-bis(1,1-dimethylethyl)-8-hydroxy-4,11-dioxo-9-(phenylmethyl)-6-[[4-(2-pyridinyl)phenyl] methyl]-2,5,6,10,13-penta aza tetra deca-nedioic acid dimethyl ester which is inhibitor of HIV- 1 protease1. Ritonavir (RTV), 2, 4, 7, 12-Tetraazatridecan-13-oic acid, 10-hydroxy-2-methyl-5-(1-methylethyl)-1-[2-(1-methylethyl)-4-thiazolyl]-3,6-dioxo-8,11-bis(phenylmethyl)-5-thiazolylmethyl ester [5S-(5R*,8R*,10R*,11R*)] is a potent cytochrcome P -450 (CYP)3A inhibitor and usually used as pharmacokinetic booster for other protease inhibitor including Atazanavir, thereby providing increased plasma level of Atazanavir2.
Literature survey reveals High Performance Liquid Chromatographic (HPLC)3-9, LC-MS10 and Ultra Performance Liquid Chromatography (UPLC)11 methods for determination of ATV as single and in combination with other drugs in human plasma. Also Spectrophotometric method for degradation studies of ATV in dosage form has been also reported12.
Analytical methods reported for RTV includes HPLC13-15, LC-MS16, Densiometry17,18 and spectrophotometry19 either as single or in combination with other drugs.
No reports were found for stability-indicating HPLC method for simultaneous determination of ATV and RTV in tablet dosage form. This paper describes simple, precise, accurate and sensitive HPLC method development and validation as well as stability study (hydrolysis, oxidation, photo-degradation and thermal degradation) as per International Conference on Harmonisation Guidelines20, 21.
MATERIALS AND METHODS:
Chemicals and Reagents:
Analytically pure samples of RTV and ATV were kindly supplied by Emcure Pharma Pvt. Ltd. (Pune, India) and Cipla Pvt. Ltd. (Kurkumbh, India) respectively. The pharmaceutical dosage form used in this study was Sinthivan tablets (Cipla Pvt. Ltd., Patalganga, India) labeled to contain 300 mg of ATV and 100 mg of RTV were procured from the local market. Methanol (HPLC grade), Potassium dihydrogen phosphate (AR grade) were purchased from Merck specialties Pvt. Ltd. (Mumbai, India).
Chromatographic Conditions:
HPLC system used was JASCO system equipped with Model PU 2080 Plus pump, Rheodyne sample injection port (20 μl), MD 2010 PDA detector and Borwin- PDA software (version 1.5). A chromatographic column Grace C18 (150 x 4.6 mm i.d. 3µm) was used. Separation was carried out at flow rate of 1 mL min-1 using 10 mM potassium dihydrogen phosphate buffer: acetonitrile: methanol (40:40:20, v/v/v) and detection at 241 nm.
Standard stock solution of ATV and RTV were prepared separately by dissolving 10 mg of each drug in 10 mL methanol to get concentration of 1000 µg mL-1.
Preparation of sample solution (Tablet Formulation Analysis):
Twenty tablets were weighed accurately and powdered. A quantity of tablet powder equivalent to 5 mg of RTV (15 mg of ATV) was weighed and transferred to 10 mL volumetric flask containing about 6 mL of water and ultrasonicated for 10 min and volume was made upto the mark with the Methanol. The solution was filtered through Whatman paper No. 41. One mL of this solution was transferred to 10 mL calibrated volumetric flask and volume was made up to the mark with the methanol to get solution of concentration 50 µg mL-1 for RTV and 150 µg mL-1 for ATV. After setting the chromatographic conditions, the tablet sample solution was injected, chromatogram was obtained and the peak areas were recorded. The injections were repeated six times and the amount of each drug present per tablet was estimated from the respective calibration curves. The % assay was found to be 100.42 ± 0.67 for ATV and 100.01 ± 0.25 for RTV (mean ± S.D.).
The system suitability was assessed by six replicate injections of the mixture containing 50 µg mL-1 and 25 µg mL-1 of ATV and RTV respectively. The resolution, peak asymmetry, number of theoretical plates and height equivalent to theoretical plate (HETP) were calculated. The values obtained demonstrated the suitability of the system for the analysis of these drugs in combination. The results obtained are represented in Table 1.
Table 1: System suitability parameters for proposed RP-HPLC method
|
Sr. No. |
Parameters |
ATV |
RTV |
|
1 |
Theoretical plates |
3411.60 |
3048.70 |
|
2 |
HETP (cm) |
0.0043 |
0.0049 |
|
3 |
Resolution |
|
3.01 |
|
4 |
Asymmetry factor |
0.88 |
0.84 |
a With respect to previous peak.
Stress degradation studies of bulk drug:
Forced degradation studies were carried out to provide evidence on how stability of drug varies under the influence of variety of environmental conditions like hydrolysis, oxidation, temperature, etc. and to establish specific storage conditions, shelf-life and retest period.
Acid treatment:
3 mL methanolic stock solution of ATV was mixed with 1 mL of 0.1 N HCl in 10 mL volumetric flasks to obtain final concentration 300 µg mL-1. The mixture was refluxed for half an hour and injected on the system. RTV was treated in similar manner to ATV.
Alkaline treatment:
3 mL methanolic stock solution of ATV was mixed with 1 mL of 0.1 N NaOH in 10 mL volumetric flasks to obtain final concentration 300 µg mL-1. The mixture was refluxed for two hours and injected on the system. RTV was treated in similar manner to ATV.
Oxidation degradation:
3 mL methanolic stock solution of ATV was mixed with 1 mL of 30% H2O2 in 10 mL volumetric flasks to obtain final concentration 300 µg mL-1. The mixture was refluxed for three hours and injected on the system. RTV was treated in similar manner to ATV.
Neutral Hydrolysis:
3 mL methanolic stock solution of ATV was mixed with 1 mL of distilled water in 10 mL volumetric flasks to obtain final concentration 300 µg mL-1. The mixture was refluxed for one hour and injected on the system. RTV was treated in similar manner to ATV.
Photo-degradation:
Photolytic studies were carried out by exposure of drug to UV light up to 200 watt hours square meter-1 and subsequently to fluorescence light illumination not less than 1.2 million lux hours. Sample was weighed, dissolved in methanol to get concentration of 300 µg mL-1 and injected on the system. RTV was treated in similar manner to ATV.
Degradation under dry heat:
Dry heat study was performed by keeping ATV in oven at 60ŗC for 7 days. A sample was withdrawn at appropriate times, weighed and dissolved in methanol to get solution of 300 µg mL-1 and injected on the system. RTV was treated in similar manner to ATV.
RESULTS AND DISCUSSION:
Optimization of chromatographic conditions:
The primary target in developing this stability indicating HPLC method is to achieve the resolution between ATV, RTV and its degradation products. To achieve the separation, we used a stationary phase C-18 column and mobile phase 10 mM potassium dihydrogen phosphate buffer, acetonitrile and methanol in ratio 40:40:20, v/v/v. The tailing factor obtained was less than two and retention time was 5.54 ± 0.02 for ATV and 6.82 ± 0.07 min. for RTV, respectively. The representative chromatogram of the standard solution of mixture is shown in Figure 1.
Result of forced degradation studies:
Forced degradation study showed the method is highly specific and no degradation products were eluted at retention time of drugs.
Figure 1: Representative chromatogram of standard mixture of ATV (50 µg mL-1, 5.54 min) and RTV (25 µg mL-1, 6.82 min)
Acid treatment:
12 % of ATV and 50 % of RTV degraded in acid condition when refluxed for half an hour. However no additional degradation peaks were seen in the chromatogram of ATV.
(a)
(b)
Figure 2: Representative chromatogram oobtained after acid treatment (a) ATV and (b) RTV with degradation product at RT 1.69 min (d1), 2.33 min (d2), 3.82 min (d3), 4.12 min (d4) and 5.00 min (d5)
Alkali treatment
56 % of degradation of RTV was observed in alkaline condition while 4 % degradation of ATV was observed with no additional degradation peak.
(a)
(b)
Figure 3: Representative chromatogram obtained after alkali treatment a) ATV (b) RTV with degradation product at RT 2.10 min (d1)
Oxidative degradation:
20 % degradation was observed for ATV when treated with 30 % H2O2 while RTV exhibited 5 % degradation.
(a)
(b)
Figure 4: Representative oxidative degradation chromatogram of (a) ATV with degradation product at RT 2.50 min (d1), 2.74 min (d2), 4.58 min (d3), 5.30 min (d4) (b) RTV with degradation product at RT 3.29 min (d1) and 3.70 min (d2)
Neutral Hydrolysis:
ATV was found to be stable in neutral condition when refluxed for one hour while 12 % of degradation was observed for RTV.
(a)
(b)
Figure 5: Representative chromatogram of neutral hydrolytic degradation (a) ATV (b) RTV with degradation product at RT 3.53 min (d1) and 4.04 min (d2)
Photo Degradation Studies:
ATV and RTV were found to be stable in ultraviolet light as well as fluorescence light.
Dry Degradation Studies:
The solid state studies showed that the drug substances ATV and RTV were stable to the effect of temperature when the powdered drug substance was exposed to dry heat at 60ŗ C for 7 days.
Peak purity results greater than 990 indicate that ATV and RTV peaks are homogeneous in all stress conditions tested. The unaffected assay of ATV and RTV in the tablet confirms the stability indicating power of the method. The forced degradation studies data are summarized in Table 2.
Table 2: Data of forced degradation studies of ATV and RTV
|
Stress conditions/ duration
|
ATV |
RTV |
|
|
(%) Degradation |
(%) Degradation |
||
|
Acidic / 0.1 N HCl/ Reflux half an hour Alkaline /0.1 N NaOH/ Reflux for 2 hours Oxidative /30 % H2O2 / Reflux for 3 hours Neutral/ H2O/ Reflux for 1 hour Photolysis UV/200 watt hours/square meter Fluorescence / 1.2 million lux hours Dry heat/ 60ŗC/ 7 days
|
12
04
20
0
0 0
0 |
50
56
05
12
0 0
0 |
Method Validation:
The method was validated for linearity, accuracy and intra-day and inter-day precision, specificity and robustness, in accordance with ICH guidelines20.
The linearity of the responses of the drugs was verified at six concentration levels, ranging from 50-500 µg mL-1 for ATV and 25- 400 µg mL-1 for RTV, respectively. The calibration graph was obtained by plotting peak area versus the concentration and data was treated by least-squares linear regression analysis. The equation of the calibration curve found for ATV y = 20224x + 13998 and for RTV y = 14636x + 48605, respectively. The calibration graphs were found to be linear in the plotted concentrations with coefficient of correlation 0.998 for ATV and 0.999 for RTV.
One set of three different concentrations of mixed standard solutions of ATV and RTV were prepared. All the solutions were analyzed thrice, in order to record any intraday variations in the results. Intra-day variation, as RSD (%), was found to be in the range of 0.33 to 1.31 for ATV and 0.58 to 1.18 for RTV. For Inter day variations study three different concentrations of the mixed standard solutions in linearity range were analyzed on three consecutive days. Interday variation, as RSD (%) was found to be in the range of 0.55 to 1.17 for ATV and 0.13 to 0.75 for RTV.
Limit of detection (LOD) and Limit of quantitation (LOQ):
LOD and LOQ were calculated as 3.3 σ/S and 10 σ/S, respectively; where σ is the standard deviation of the response (y-intercept) and S is the slope of the calibration plot. The LOD of ATV and RTV were found 1.73 µg mL-1 and 5.54 µg mL-1, respectively. The LOQ of ATV and RTV were 5.24 µg mL-1 and 16.80 µg mL-1, respectively.
Accuracy:
To check the accuracy of the method, recovery studies were carried out by addition of standard drug solution to pre-analyzed sample solution at three different levels 80 %, 100 % and 120 %. Basic concentration of sample chosen was 150 µg mL-1 of ATV and 50 µg mL-1 of RTV from tablet solution. The drug concentrations were calculated from respective linearity equation. The results obtained are shown in Table 3.
Specificity:
The specificity of the method was ascertained by peak purity profiling studies. The peak purity values were found to be more than 991, indicating the no interference of any other peak of degradation product, impurity or matrix.
Robustness of the method was determined by carrying out the analysis under conditions during which flow rate of the mobile phase and wavelength at which the drugs were recorded were altered and the effect on the area of drug were noted. It was observed that there were no marked changes in the chromatograms, which demonstrated that the RP-HPLC method developed is robust.
Conclusion:
The developed method is stability indicating and can be used for assessing the stability of ATV and RTV in bulk drug and pharmaceutical dosage form. The developed method is accurate, precise, specific and robust.
AcknowledgementS:
The authors express their gratitude to Cipla Pvt. Ltd. (Kurkumbh, India) and Emcure Pharmaceuticals (Pune, India) for the gift sample of pure ATV and RTV respectively.
Table 3: Recovery Studies of ATV and RTV
|
Drug |
Amount taken (mg mL-1) |
Amount added (mg mL-1) |
Total amount found (mg mL-1) |
% Recovery |
% RSDa |
|
ATV |
150 |
120 |
270.10 |
100.03 |
0.83 |
|
150 |
150 |
299.11 |
99.70 |
0.54 |
|
|
150 |
180 |
332.90 |
100.88 |
0.13 |
|
|
RTV |
50 |
40 |
90.55 |
100.62 |
0.96 |
|
50 |
50 |
99.59 |
99.59 |
0.07 |
|
|
50 |
60 |
109.11 |
99.19 |
0.22 |
*Average of three determinations, R.S.D. is relative standard deviation
1. http://en.wikipedia.org/wiki/Atazanavir (accessed on 01/03/2012)
2. http://www.sinoapi.com/Pharmacopoeia/pharmacopoeia-usp30.asp?cas=155213-67-5 (accessed on 01/03/2012)
3. Srinivasu K, Venkateswara Rao J, Appala Raju N and Mukkanti K. A validated RP-HPLC method for the determination of atazanavir in pharmaceutical dosage form. E-Journal of Chemistry 8; 2011: 453-456.
4. Colombo S, Guignard N, Marzolini C, Telenti A, Biollaz J and Decosterd LA. Determination of the new HIV-protease inhibitor atazanavir by liquid chromatography after solid-phase extraction. J. Chromatogr. B 8109; 2004: 25-34.
5. Padmalatha M, Vanitha PK and Eranna D. Validated reversed phase high performance liquid chromatography method for the estimation of atazanavir sulfate in pharmaceutical formulations. Oriental J. Chem. 26; 2010: 123-127.
6. Loregian A, Pagni S, Ballarin E, Sinigalia E, Parisi SG and Palł G. Simple determination of the HIV protease inhibitor atazanavir in human plasma by high-performance liquid chromatography with UV detection. J. Pharm. Biomed. Anal. 42; 2006: 500-505.
7. Müller AC and Kanfer I. An efficient HPLC method for the quantitative determination of atazanavir in human plasma suitable for bioequivalence and pharmacokinetic studies in healthy human subjects. J. Pharm. Biomed. Anal. 53; 2010: 113-118.
8. Takahashi M, Yoshida M, Oki T, Okumura N, Suzuki T and Kaneda T. Conventional HPLC method used for simultaneous determination of the seven HIV protease inhibitors and nonnucleoside reverse transcription inhibitor efavirenz in human plasma. Bio. Pharm. Bull. 28; 2005: 1286-1290.
9. Else L, Watson V, Tjia J, Hughes A, Siccardi M and Khoo S. Validation of a rapid and sensitive high-performance liquid chromatography-tandem mass spectrometry assay for the simultaneous determination of existing and new antiretroviral compounds. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 878; 2010: 1455-1465.
10. Seshachalam U, Narasimha Rao DVL, Haribabu B, Chandrasekhar KB. Determination of atazanavir in the presence of its degradation products by a stability-indicating LC method. Chromatographia 65; 2007: 355-358.
11. Antunes MV, Poeta J, .Ribeiro JP, Sprinzb E and Linden R. Ultra-performance liquid chromatographic method for simultaneous quantification of HIV non-nucleoside reverse transcriptase inhibitors and protease inhibitors in human plasma. J. Braz. Chem. Soc. 1; 2010: 1-8.
12. Dey S, Reddy YV, Reddy T, Sahoo SK, Murthy PN, Mohapatra S and Patro SS. Method development and validation for the estimation of atazanavir in bulk and pharmaceutical dosage forms and its stress degradation studies using UV-vis spectrophotometric method. Int. J. Pharm. Bio. Sci. 1; 2010.
13. Suneetha A, Kathirvel S Ramachandrik G. Validated RP- HPLC method for simultaneous estimation of lopinavir and ritonavir in combined dosage form. Int. J. Pharm. Pharm. Sci. 3; 2011: 4951.
14. Phechkrajang CM, Thin EE, Sratthaphut L, Nacapricha D and Wilairat P. Quantitative determination of lopinavir and ritonavir in syrup preparation by liquid chromatography. J. Pharm. Sci. 36; 2009: 1-12.
15. Temghare GA, Shetye SS and Joshi SS. Rapid and sensitive method for quantitative determination of lopinavir and ritonavir in human plasma by liquid chromatography- tandem mass spectrometry. E-Journal of Chemistry 6; 2009: 223-230.
16. Donato EM, Dias CL, Rossi RC, Valente RS, Froehlich PE and. Bergold AM. LC method for studies on the stability of lopinavir and ritonavir in soft gelatin capsules. Chromatographia 63; 2006: 437-443.
17. Sulebhavikar AV, Pawar UD, Mangoankar KV and Prabhu ND. HPTLC method for simultaneous determination of lopinavir and ritonavir in capsule dosage form. E-Journal of Chemistry 5; 2008: 706-712.
18. Patel GF, Vekariya NR and Bhatt HS. Application of TLC-densitometry method for simultaneous determination of lopinavir and ritonavir in capsule dosage form. Oriental J. Chem. 25; 2009: 727-730.
19. Nagulwar VP and Bhusari KP. Simultaneous estimation of ritonavir and lopinavir by absorption ratio UV pectrophotometric method in combined tablet dosage form. Der Pharmacia Lettrers 2; 2010: 196-200.
20. International Conference on Harmonization (2005) ICH harmonised tripartite guideline Validation of analytical procedures: text and methodology Q2 (R1) ICH, Geneva, Nov (2005).
21. International Conference on Harmonization (ICH), Stability testing of new drug substances and products, Q1A (R2), (2003).
Received on 24.09.2014 Modified on 07.10.2014
Accepted on 15.10.2014 © RJPT All right reserved
Research J. Pharm. and Tech. 7(11): Nov. 2014 Page 1231-1236